Narda 3502/02
High-frequency field strength measurements for E & H fields
Technical highlights
- Frequency range: 200 MHz – 6 GHz
- Measurement principle: Three-axis, isotropic, direction-independent E-field measurement
- Compatibility: For the Narda SRM‑3006 Selective Radiation Meter
User highlights
- ✓ Quick measurement
- ✓ Future-proof
- ✓ Investment protection
Applications:
- Telecommunications
- Military and Security
- Testing laboratories and certification bodies
- Semiconductor and Electronics Development
- Research and universities
The Narda Three Axis E Field Antenna 3502 02 is an isotropic measurement antenna for the precise measurement of electric fields in the radio frequency range. It measures simultaneously in three spatial directions and enables the field strength to be determined independently of direction.
The three-axis measurement ensures that the field strength is measured accurately, regardless of the antenna’s position. This is particularly important in complex electromagnetic environments where fields originate from different directions.
A key advantage of the Narda Three Axis E Field Antenna 3502 02 lies in its high measurement accuracy and reproducibility. It delivers consistent results and enables precise assessments of EMF exposure in a variety of applications.
The antenna is compatible with standard Narda measuring instruments and can be easily integrated into existing systems. This makes it suitable for both mobile measurements and fixed installations.
Typical areas of application include:
• Measurement of electric fields in the RF range
• EMF assessments in the workplace
• Use by public authorities and network operators
• Research and development
The Narda Three Axis E Field Antenna 3502 02 enables reliable and comprehensive measurement of electromagnetic fields. It facilitates precise analysis and helps ensure compliance with relevant standards.
For professional EMF measurements where accuracy and independence of orientation are crucial, this antenna offers a high-performance solution.
Technical properties and features of this product.
| Spezifikation | Wert |
|---|---|
| Frequency range | 200 MHz – 6 GHz (optional older version: 27 MHz – 3 GHz) |
| Antenna type | three-axis isotropic E-field antenna |
| Measurement principle | Frequency-selective EMF measurement with the SRM-3006 |
| Dynamic range | approx. 0.2 mV/m – 200 V/m |
| Compatibility | Compatible with the Narda SRM‑3006 |
| Directional characteristics | isotropic (independent of direction) |
| Application | EMF measurements in mobile communications, broadcasting and industrial environments |
| Accuracy/Performance | High precision and an extended frequency range compared with previous models |
| Special features | automatic detection of calibration data within the system |
| Calibration data | Stored internally / automatic detection by SRM-3006 |
Download the product documents directly or request further documents.
Request further documents